HUATEC GROUP CORPORATION

HUATEC GROUP CORPORATION

Manufacturer from China
Verified Supplier
1 Years
Home / Products / Magnetic Particle Testing /

Particle Size Analysis Magnetic Particle Testing Equipment EN ISO 9934-2

Contact Now
HUATEC GROUP CORPORATION
Visit Website
City:beijing
Province/State:beijing
Country/Region:china
Contact Person:MrNoah
Contact Now

Particle Size Analysis Magnetic Particle Testing Equipment EN ISO 9934-2

Ask Latest Price
Video Channel
Brand Name :HUATEC
Model Number :MTU
Certification :ISO, CE, GOST
Place of Origin :Beijing
MOQ :1 set
Supply Ability :500pcs per month
Packaging Details :Standard export package
Standard :EN ISO 9934-2
Power :Electronic
Usage :magnetic testing equipment
Application :Particle size distribution classification
Warranty :12 Months
Function :Particle size analysis
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

MTU test block magnetic particle detector magnaflux magnetic particle test equipment

Instructions
Reference Block Type 1 (MTU Nr.3)


Description:
The reference block type 1 is a residually magnetized metal disc, where, caused by a special treatment, a network of coarse and fine cracks had been generated. The magnetic field of the reference block generates stray fluxes of different strength across these cracks. More sensitive
inspection media result in a more distinct indications as compared to the indications when lower
sensitivity is used. The central bore hole is made for easier handling.


Use:
The reference block type 1 according to EN ISO 9934-2 (former description "MTU test block") isused for checking and supervision of the indication sensitivity of magnetic particle crack detection agents. The surface of the residually magnetized reference block contains a network of coarse and fine cracks. After dipping or spraying with the magnetic particle solution, the crack indications are
evaluated.


Application:
The reference block type 1 is immersed or carefully sprayed with the well-mixed inspection medium.Redundant inspection medium should be drained off. Then the crack indications are to be viewed under appropriate lighting conditions. You should then inspect the weaker indications as this is where differences in particle sensitivity becomes obvious.


Note:
The crack indication of the reference block type 1 may be affected by contamination, strong magnetic fields, or mechanical influences. To give correct indications, the reference block should be free of dirt, oil, grease and magnetic particles from former applications. Therefore, the reference
block should be cleaned after each application (wipe off with a wet cloth). The reference block should not be exposed to strong magnetic fields such as from MP crack detectors, magnetizing and demagnetizing coils, magnetic holding plates, etc. as they may influence the residual field of the block by altering the field strength and direction.